User manual

page 20 of 25 Report No. 55S072148/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
20 TABLE: dielectric strength P
Insulation under test
(area from insulation
diagram)
Insulation type: (OP-operational /
BI-basic / SI-supplementary /
DI-double / RI-reinforced)
Reference
voltage (V)
Test
voltage
(V)
Remarks
Before humidity
A-a2
Primary to LCD
Enclosure
DI-double 250 4852 Vac
Pass
A-e
Primary to LCD SIP/
SOP
DI-double 354 4852 Vac
Pass
After humidity
A-a2
Primary to LCD
Enclosure
DI-double 250 4852 Vac
Pass
A-e
Primary to LCD SIP/
SOP
DI-double 354 4852 Vac
Pass
Supplementary information:
21 TABLE: mechanical strength P
Part under test Test (impact, drop, force, handle, rough
handling, mobile)
Remarks
Metal enclosure top/ side/ rear Force Test: 45N Pass
Metal enclosure top/ side/ rear Impact Test by the impact hammer 0.5 J Pass
LCD Monitor with base Drop Test at height: 5 cm Pass
Supplementary information:
24 TABLE: - stability P
Part under test Test condition Remarks
LCD Monitor
10 tilt conducted on front, rear and sides
Pass, not overbalance
Supplementary information: