User manual

page 11 of 25 Report No. 55S072148/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
d) Environmental conditions for transport and
storage specified in accompanying documents
and marked on packaging
Ambient: -20
o
C to +60
o
C
Humidity: 10% to 90%
Air Pressure: 187 to 1060hPa
was provided in the user
manual.
P
16 ENCLOSURES AND PROTECTIVE COVERS
P
16a Equipment enclosed to protect against contact
with live parts, and with parts which can become
live (finger, pin, hook test)
In compliance with the leakage
current requirements.
(see appended table)
P
Insertion or removal of lamps - protection against
contact with live parts provided
No lamps inside the
equipment.
N
16b Opening in a top cover positioned that
accessibility of live parts by a test rod is prevented
Components enclosed in the
EUT cannot be touched by the
test rod.
P
16c Conductive parts accessible after the removal of handles, knobs, levers N
- have a resistance of not more than 0.2
No such parts. N
- separated from live parts by one of the means
described in Sub-clause 17g
N
16d Parts with voltage exceeding 25V a.c. or 60V d.c.
which cannot be disconnected by external mains
switch or plug protected against contact
No accessible internal
components.
N
16e Removable enclosures protecting against contact with live parts P
- Removal possible only with the aid of a tool The removal of the enclosure
need the aid of tool.
P
- Use of automatic device making parts not live
when the enclosure is opened or removed
N
- Exception 16e applied to the following parts ....... : N
16f Openings for the adjustment of controls using a
tool. The tool not able to touch basic insulation or
any live parts
No pre-set control which may
be adjusted during normal
used.
P