User manual

page 8 of 25 Report No. 55S072148/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
k) For permanently connected devices the supply
connections are clearly marked adjacent to the
terminals (or in accompanying documents for small
equipment)
Not permanently connected
equipment.
N
l) Statement for suitable wiring materials at
temperatures over 75 °C
Not permanently connected
equipment.
N
n) Capacitors and/or circuit parts marked as
required in Sub-clause 15c
No such devices. N
6.3 Marking of controls and instruments P
a) Mains switch clearly identified No mains switch. N
- ON and OFF positions marked according to
Symbols 15 and 16 of table D1 or indicated by an
adjacent indicator light
N
b) Indication of different positions of control
devices and switches
Indication of control devices
were provided on the
equipment.
P
c) Indication of the direction in which the magnitude
of the function changes, or an indicating device
N
f) The functions of operator controls and indicators
are identified
Provided in the user manual. P
g) Numeric indications of parameters are in SI
units except for units listed in Am. 2
Provided in SI units. P
6.4 Symbols P
Used symbols comply with Appendix D or IEC 417
and/or IEC 878 or ISO publications (if applicable)
The stand-by symbol is
marked with the correct
symbol (60417-1-IEC-5009).
P
6.5 Colors of the insulation of conductors N
a) Protective earth conductor has green/yellow
insulation
N
b) All insulations of internal protective earth
conductors are green/yellow at least at their
terminations
Evaluated in the certified
power adapter.
N
c) Only protective or functional earthing, or potential
equalization conductors are green/yellow
Evaluated in the certified
power adapter.
N
d) Color of neutral conductor .................................... : N
e) Colors of phase conductor(s) .............................. : N
- Compliance with IEC 227 and IEC 245 N
f) Additional protective earthing in multi-conductor,
cords are marked green/yellow at the ends of the
additional conductors
N
6.6 Medical gas cylinders and connections N