User manual

page 6 of 25 Report No. 55S072148/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
6 IDENTIFICATION, MARKING AND DOCUMENTS
P
6.1 Marking on the outside of equipment or equipment parts P
c) Markings of the specific power supply affixed
No applicable. No specific
power supply affixed.
N
d) If marking is not practicable due to size or nature
of enclosure, information is included in
accompanying documents
P
e) Name and/or trademark of the manufacturer or
supplier ..................................................................... :
DIVA Laboratories Ltd. or
DIVA
P
f) Model or type reference ........................................ :
GOD971xM, GOK971xM,
ER-192
P
g) Rated supply voltages or voltage range(s)
24V
P
Number of phases .................................................... : N
Type of current ......................................................... : DC Symbol 417-5031 is used. P
h) Rated frequency or rated frequency range(s)
(Hz) .......................................................................... :
N
j) Rated power input (VA, W or A)............................ : 3.0A P
k) Power output of auxiliary mains socket-outlets No socket-outlet. N
l) Class II symbol Class I equipment. N
Symbol for degree of protection against ingress of
water provided .......................................................... :
IPX2, wording in manual. P
Symbol for protection against electric shock ........... : No applied part. N
If equipment has more than one applied part with
different degrees of protection, the relevant
symbols are clearly marked on such applied parts,
or on or near relevant outlets
N
Symbol for protection of defibrillation-proof applied
parts
N
Symbol 14 from Table DI for defibrillation-proof with
protection partly in patient cable
N
m) Mode of operation (if no marking, suitable for
continuous operation)
Continuous operation. N
n) Types and rating of external accessible fuses .... : No external fuses. N
p) Ratings of external output ................................... : Not applicable. N
q) Symbol for physiological effect(s): P
- attention, consult accompanying documents Symbol 14 of Table DI was
used in the User Manual.
P
- non-ionizing radiation, or symbols as adopted by
ISO or IEC 417
N