Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
- non-ionizing radiation, or symbols as adopted by
ISO or IEC 417
N
r) Anaesthetic-proof symbol: AP or APG ................. : Not of category AP or APG. N
s) Dangerous voltage symbol No accessible high voltage
terminal devices.
N
t) Special cooling requirements No such devices. N
u) Limited mechanical stability The equipment did no
overbalance.
P
v) Protective packing requirement(s) No special measure was taken
during transportation or
storage.
N
- Marking(s) for unpacking safety hazard(s) N
- Equipment or accessories supplied sterile,
marked as sterile
N
y) Potential equalization terminal N
- Functional earth terminal N
z) Removable protective means No such devices. N
Durability of marking test P
6.2 Marking on the inside of equipment or equipment parts P
a) Nominal voltage of permanently installed
equipment
Not permanently connected
equipment.
N
b) Maximum power loading for heating elements or
holders for heating lamps
No heating elements or lamps
inside the equipment.
N
c) Dangerous voltage symbol Provided on the inverter,
located inside the equipment.
P
d) Type of battery and mode of insertion No battery inside the
equipment.
N
- Marking referring to accompanying documents
used for battery not intended to be changed by the
operator
N
e) Fuses accessible with a tool identified either by
type and rating or by a reference to diagram
Fuses were not accessible. N
f) Protective earth terminal Appliance inlet used. P
g) Functional earth terminal No functional earth terminal. N
h) Supply neutral conductor in permanently
installed equipment (N)
Not permanently installed
equipment.
N
j) Markings required in 6.2 f), h), k) ,and l) remain
visible after connection and are not affixed to parts
which have to be removed
N