User manual

page 6 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
6 Identification, marking and documents P
6.1 Marking on the outside of equipment or equipment parts P
c) Markings of the specific power supply affixed
No applicable. No specific
power supply affixed.
N
d) If marking is not practicable due to size or nature
of enclosure, information is included in
accompanying documents
Not applicable.
N
e) Name and/or trademark of the manufacturer or
supplier ..................................................................... :
DIVA Laboratories Ltd. or
DIVA
P
f) Model or type reference ........................................ :
GODx7yzMI, GOKx7yzMI.
ER-191
(see Remark 4 on page 4)
P
g) Rated supply voltages or voltage range(s)
100-240 V~
P
Number of phases .................................................... :
Single phase.
N
Type of current ......................................................... :
AC
P
h) Rated frequency or rated frequency range(s)
(Hz) .......................................................................... :
50/60 Hz
P
j) Rated power input (VA, W or A) ............................ : 1.2-0.6 A P
k) Power output of auxiliary mains socket-outlets No socket-outlet. N
l) Class II symbol Class I equipment. N
Symbol for degree of protection against ingress of
water provided .......................................................... :
IPX2, wording in manual. P
Symbol for protection against electric shock ........... : No applied part. N
If equipment has more than one applied part with
different degrees of protection, the relevant
symbols are clearly marked on such applied parts,
or on or near relevant outlets
N
Symbol for protection of defibrillation-proof applied
parts
No defibrillation applied parts. N
Symbol 14 from Table DI for defibrillation-proof with
protection partly in patient cable
N
m) Mode of operation (if no marking, suitable for
continuous operation)
Continuous operation. N
n) Types and rating of external accessible fuses .... : No external fuses. N
p) Ratings of external output ................................... : Not applicable. N
q) Symbol for physiological effect(s): N
- attention, consult accompanying documents Symbol 14 of Table DI was
used in the User Manual.
P