User manual
page 5 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
5 CLASSIFICATION P
5.1 Type of protection against electric shock P
Class I equipment Class I equipment. P
Class II equipment Class I equipment. N
Internally powered equipment Class I equipment. N
5.2 Degree of protection against electric shock N
Type B applied part
No applied part.
N
Type BF applied part N
Type CF applied part N
Not classified - no applied parts
As the equipment is not
intended to be connected to the
patient and does not have any
patient applied parts, it is
should not be marked with the
type B applied part symbol.
Nevertheless, the product
complies with the requirements
for type B applied part
concerning protection against
electric shock.
P
5.3 Classification according to the degree of protection
against ingress of water as detailed in the current
edition of IEC 529 (see 6.1.1) ..................................
::
IPX2, only for model: ER-191.
(see separate test report)
P
5.4 Methods of sterilization or desinfection
Not applicable.
N
5.5 Equipment not suitable for use in the presence of
flammable mixtures
The equipment is not AP or
APG category equipment.
N
Category AP equipment N
Category APG equipment N
5.6 Mode of operation: P
-continuous operation
Continuous operation.
P
-short-time operation, specified operation; period... : —
-intermittent operation, specified operation; rest
period ....................................................................... :
—
-continuous operation with short-time, stated
permissible loading time ..........................................
::
—
-continuous operation with intermittent, stated
permissible loading/rest time ................................... :
—