page 18 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
a) Thermal cut-outs which have to be reset by a
soldering not fitted in equipment
No thermal cut-out inside the
equipment.
N
Thermal safety devices provided where necessary
to prevent operating temperatures exceeding the
limits
No thermal safety devices
inside the equipment.
N
Independent non-self-resetting thermal cut-out
provided where a failure of a thermostat could
constitute a safety hazard
No such device. N
Audible warning provided where the loss of
function caused by operation of a thermal cut-out
presents a safety hazard
N
Self-resetting thermal cut-outs and self-resetting
over-current releases operated 200 times
No such device. N
Non-self resetting over-current releases operated
10 times
N
56.6b Thermostats with varying temperature settings
clearly indicated
No thermostat inside the
equipment.
N
Operating temperature of thermal cut-outs
indicated
N
56.7 Batteries N
a) Battery compartments: N
- adequately ventilated No battery inside the
equipment.
N
- accidentally short-circuiting is prevented N
b) Incorrect polarity of connection prevented N
56.8 Indicators - unless indication provided by other means (from the normal operation
position), indicator lights are used (color see 6.7):
P
- to indicate that equipment is energized N
- to indicate the operation of non-luminous
heaters if a safety hazard could result
No heaters inside the
equipment.
N
- to indicate when output exists if a safety hazard
could result
N
- charging mode indicator provided No battery inside the
equipment.
N
56.10 Actuating parts of controls No actuating parts of controls
used.
N
56.10b Actuating parts are adequately secured to prevent
them from working loose during normal use
N
Controls are secured to prevent the movement
relative to scale marking (safety related only)
N