page 17 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
56 COMPONENTS AND GENERAL ASSEMBLY
P
List of critical components (see appended table 56.1) P
56.1b Ratings of components not in conflict with the
conditions of use in equipment
The components are used
according to their rating.
P
Ratings of mains components are identified The critical components were
used within the intended
rating.
P
56.1d Components, movements of which could result in
a safety hazard mounted securely
The movement of components
is prevented.
P
56.1f Conductors and connectors secured and/or
insulated to prevent accidental detachment
resulting in a safety hazard
No connection was likely to be
detached.
P
56.3a Connectors provide separation required by
Sub-clause 17g
Double or reinforce insulation
provided.
P
Plugs for connection of patient circuit leads can
not be connected to other outlets on the same
equipment
No connection to the patient. N
Medical gas connections not interchangeable N
56.3b Accessible metal parts can not become live when
detachable interconnection cord between different
parts of equipment is loosened or broken
Interconnection cord only
connected to Safety Extra Low
Voltage.
P
56.3c Leads with conductive connection to a patient are
constructed such that no conductive connection
remote from the patient can contact earth or
hazardous voltages.
No patient connection. N
56.4 Connections of capacitors P
Not connected between live parts and non-
protectively earthed accessible parts
P
If connected between mains part and protectively
earthed metal parts comply with:
IEC Publication 384-14
Evaluated in the certified
power supply.
P
Enclosure of capacitors connected to mains part
and providing only basic insulation, is not secured
to non-protectively earthed metal parts
No such device. N
Capacitors or other spark-suppression devices
are not connected between contacts of thermal
cut-outs
No thermal cut-off used. N
56.5 Protective devices which cause disconnection
from the supply mains by producing a short-circuit
not provided in equipment
No such device. N
56.6 Temperature and overload control devices N