page 15 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
44 OVERFLOW, SPILLAGE, LEAKAGE, HUMIDITY, INGRESS OF LIQUIDS,
CLEANING, STERILIZATION AND DISINFECTION
P
44.2 Equipment contain a liquid reservoir: N
- the equipment is electrically safe after 15% overfill
steadily over a period of 1 min
No liquid reservoir inside the
equipment.
N
- transportable equipment is electrically safe after
additionally having been tilted through an angle of
15º in the least favorable direction(s) (if necessary
with refilling)
N
44.3 Electrical properties of the equipment do not
change in connection of spillage test (200 ml of
water)
No use of liquid inside the
equipment.
N
44.4 Liquid which might escape in a single fault
condition does not wet parts which may cause a
safety hazard
N
44.5 Equipment sufficiently protected against the
effects of humidity
30ºC, 168 hours, 93% RH.
(See appended table)
P
44.6 Enclosures designed to give a protection against
harmful ingress of water classified according to
IEC Publication 529
IPX2, only for model: ER-191.
(see separate test report)
P
44.7 Equipment capable of withstanding cleaning,
sterilization or disinfection without deterioration of
safety provisions
Testing performed according
to method described in the
user manual.
P