User manual
page 14 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
24 STABILITY IN NORMAL USE (SEE APPENDED TABLE 24)
P
24.1 Equipment does not overbalance during normal
use when tilted through an angle of 10º
The EUT did not overbalance
at 10º.
P
24.3 Equipment overbalances when tilted through an angle of 10º N
- does not overbalance when tilted through an
angle of 5º in any position excluding transport
N
- carry a warning notice stating that transport
should only be undertaken in a certain position
N
- in the position specified for transport does not
overbalance when tilted to an angle of 10º
N
24.6a Equipment or its parts with a mass of more than 20 kg is provided with: N
- suitable handling devices (grips etc.), or The weight of the equipment,
ER-191 is 8.2 kg. (< 20 kg)
N
- instructions for lifting and handling during
assembly
N
24.6b b) On portable equipment with a mass of more
than 20 kg carrying handle(s) is (are) so situated
that equipment may be carried by 2 or more
persons
N
42 EXCESSIVE TEMPERATURES
P
42.1 Equipment does not attain temperatures
exceeding the values given in Table Xa over the
range of ambient temperatures per Clause 10.2.1
(see appended table 42) P
42.2 Equipment does not attain temperatures
exceeding the values given in Table Xb at 25ºC
ambient
(see appended table 42) P
42.3 Applied parts not intended to supply heat have
surface temperatures not exceeding 41ºC
N
42.5 Guards to prevent contact with hot surfaces
removable only with a tool