page 13 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
19 CONTINUOUS LEAKAGE CURRENTS AND PATIENT AUXILIARY CURRENTS
P
19.1b Leakage currents (see appended table 19) P
- earth leakage current P
- enclosure leakage current P
- patient leakage current No patient leakage current. N
- patient auxiliary current No patient auxiliary current. N
20 DIELECTRIC STRENGTH
P
Overall compliance with Clause 20 (see appended table 20) P
21 MECHANICAL STRENGTH
P
21a Sufficient rigidity of an enclosure tested by:
force of 45 N
(see appended table 21) P
21b Sufficient strength of an enclosure tested by:
impact hammer
No damage after the test. P
21c On portable equipment carrying handles or grips
withstand the requirements of the loading test
No handle and grip on the
enclosure.
N
21.3 No damage to parts of patient support and/or
immobilization system after the loading test
No such parts. N
21.5 Hand held equipment or equipment parts are safe
after drop test
No hand held equipment. N
21.6 Portable and mobile equipment is able to
withstand rough handling
(see appended table 21) P