User manual

page 9 of 27 Report No. 55S072147/NCH/CZ
IEC 601 + Am. 1& 2
Clause Requirement + Test Result - Remark Verdict
TRF No. I6011__C TRF originator: UL
- Compliance with IEC 227 and IEC 245 N
f) Additional protective earthing in multi-conductor,
cords are marked green/yellow at the ends of the
additional conductors
Multi-conductor cord not used. N
6.6 Medical gas cylinders and connections N
a) In accordance with ISO ISO/R 32 No such device. N
b) Identification of connection point N
6.7 Indicator lights and push-buttons N
a) Red indicator lights used exclusively to indicate
a warning of danger and/or a need for urgent
action
No red indicator or light used. N
- Yellow used to indicate caution or attention
required
N
- Yellow used to indicate caution or attention
required
N
b) Color red used only for push-buttons by which a
function is interrupted in case of emergency
N
6.8 ACCOMPANYING DOCUMENTS P
6.8.1 Equipment accompanied by documents
containing at least instructions for use, a technical
description and an address to which the user can
refer
Provided in the user manual. P
Classifications specified in Clause 5 included in
both the instructions for use and the technical
description
Provided in the user manual. P
Markings specified in Sub-clause 6.1 included in
the accompanying documents if they have not
been permanently affixed to equipment
Provided on the equipment. N
Warning statements and the explanation of
warning symbols provided in the accompanying
documents
Provided in the user manual. P
6.8.2 Instructions for use P
a) General information provided in instructions for
use
Provided in the user manual. P
- state the function and intended application of the
equipment
Provided in the user manual. P
- include an explanation of: the function of controls,
displays and signals
Provided in the user manual. P
- the sequence of operation Provided in the user manual. P
- the connection and disconnection of detachable
parts and accessories
Provided in the user manual. P