24
To Congure MPO Connector
Main Menu and Settings
Note: It is important to set the correct MPO ber conguration prior to inspection! If the MPO
connector setting is not congured correctly, the auto-analysis results will likely show “false negatives”
and possibly “false positives”.
To Congure MPO Connector
• From the Main Menu > Settings, highlight Connector
A
.
• Press Select or Right arrow button to display the Connector menu, which shows a list of the
available Connector options
B
.
• Highlight MPO option
C
, press Select.
• From the displayed list
D
, highlight the desired MPO connector, press Select.
– Note that for MPO-12 and MPO-24, you would need to select ber conguration
(
E
for MPO-12;
E1
for MPO-24), and then press Select.
• Important: Note that selected MPO conguration (
F
for MPO-12/
F1
for MPO-24) is displayed at
the bottom of Connector menu.