Understanding MPO Connector Key Orientation Settings
Understanding Fiber Numbering After The Image Capture
The goal of the MPO Key Up/Down settings is to provide users with
the correct information about ber numbering (pinout) after the
image capture.
When inspecting MPO connector using bulkheads with opposed
keys, it is important to dene the MPO key orientation in the
in Live Image mode prior to the Image Capture.
• The MPO bulkhead key position indication (UP or DOWN) is
provided in the upper right corner of the display
A
.
• At power-up, FOCIS Lightning defaults to key Up orientation.
• If needed, press the Left Arrow key
B
to toggle the MPO
connector key orientation to DOWN.
• This setting should be dened in the in Live Image mode prior to
the Image Capture.
Getting Started
After the ber end-face image is captured, FOCIS Lightning transitions to the Captured Image mode
displaying inspection results.
• Default display:
– Fiber #1 location is indicated by Blue box -
– Passing ber is indicated by Green circle -
– Failing ber is indicated by Red circle -
• When the MPO key orientation is set to Up
C
in the Live mode, ber #1 is located at top-right
D
as
indicated by Blue box.
• When the MPO key orientation is set to Down
E
,in the Live mode, after image capture, ber #1 is
located at bottom-left
F
as indicated by Blue box.
A
B
Live Image Captured Image
C
D
Key UP
Fiber #1
Live Image Captured Image
E
Key DOWN
Fiber #1
F