User manual
    AFG-2000 Series User Manual 
48 
3.  Use the arrow 
keys, scroll wheel 
and Enter key to 
edit the modulation 
depth. 
→ 
Enter
Use the keypad 
and the % key to 
enter a new depth. 
0
/
321
4
7 8
5
9
6
→ 
%
Range 
Depth 
0% ~ 120% 
Default 
100% 
Example: 
DEP= 100% 
Note 
When the modulation depth is greater than 
100%, the output cannot exceed ±5VPeak 
(50Ω load). 
If an external modulation source is 
selected, modulation depth is limited to 
±5V from the MOD input port on the rear 
panel. For example, if the modulation depth 
is set to 100%, then the maximum 
amplitude is +5V, and the minimum 
amplitude is -5V. 