User manual
    REMOTE INTERFACE 
127 
Example 
SOUR1:FREQ:STAR +2.0000E+03 
Sets the start frequency to 2kHz. 
Query Syntax 
SOURce[1]:FREQuency:STARt? [MINimum| 
MAXimum] 
Return Parameter 
<NR3> 
Returns the start frequency in Hz. 
Query Example 
SOUR1:FREQ:STAR? MAX 
>+2.0000E+07 
Returns the maximum start frequency allowed. 
SOURce[1]:FREQuency:STOP 
Set
Query
Description 
Sets the stop frequency of the sweep. 1 kHz is the 
default start frequency. 
Note 
To sweep up set the stop frequency higher than the 
start frequency. Set the stop frequency lower than 
the start frequency to sweep down. 
Syntax 
SOURce[1]:FREQuency:STOP 
{<frequency>|MINimum|MAXimum} 
Parameter 
<frequency> 
0.1Hz ~ 25MHz* 
0.1Hz ~ 1MHz (Ramp) 
*AFG-2112 limited to 12MHz, AFG-2105 limited to 
5MHz. 
Query Example 
SOUR1:FREQ:STOP +2.0000E+03 
Sets the stop frequency to 2kHz. 
Query Syntax 
SOURce[1]:FREQuency:STOP? [MINimum| 
MAXimum] 
Return Parameter 
<NR3> 
Returns the stop frequency in Hz. 
Example 
SOUR1:FREQ:STOP? MAX 
>+2.0000E+07 
Returns the maximum stop frequency allowed. 