SOURce[1]:SWEep:STATe 
Set
Query
Description 
Sets or disables Sweep mode. By default sweep is 
disabled. Sweep must be enabled before setting 
other parameters.  
Note 
Any modes will be disabled if sweep mode is 
enabled. 
Syntax 
SOURce[1]:SWEep:STATe {OFF|ON} 
Example 
SOUR1:SWE:STAT ON 
Enables sweep mode. 
Query Syntax 
SOURce[1]:SWEep:STATe? 
Return Parameter 
0 
Disabled (OFF) 
1 
Enabled (ON) 
Query Example 
SOUR1:SWE:STAT? 
>1 
Sweep mode is currently enabled. 
SOURce[1]:FREQuency:STARt 
Set
Query
Description 
Sets the start frequency of the sweep. 100Hz is the 
default start frequency. 
Note 
To sweep up set the stop frequency higher than the 
start frequency. Set the stop frequency lower than 
the start frequency to sweep down. 
Syntax 
SOURce[1]:FREQuency:STARt 
{<frequency>|MINimum|MAXimum} 
Parameter 
<frequency> 
0.1Hz ~ 25MHz* 
0.1Hz ~ 1MHz (Ramp) 
*AFG-2112 limited to 12MHz, AFG-2105 limited to 
5MHz. 