    REMOTE INTERFACE 
107 
The maximum amplitude that can be set for the 
current function is 5 volts. 
SOURce[1]:DCOffset 
Set
Query
Description 
Sets or queries the DC offset for the current mode. 
Note 
The offset parameter can be set to MINimum or 
MAXimum. The default offset is 0 volts. The offset 
is limited by the output amplitude as shown 
below. 
|Voffset| < Vmax – Vpp/2 
If the output specified is out of range, the 
maximum offset will be set. 
The maximum offset is ±5V into 50Ω).  
Syntax 
SOURce[1]:DCOffset {< offset> 
|MINimum|MAXimum} 
Example 
SOUR1:DCO MAX 
Sets the offset to the maximum for the current 
mode. 
Query Syntax 
SOURce[1]:DCOffset? {MINimum|MAXimum} 
Return Parameter 
<NR3> 
Returns the offset for the current mode. 
Query Example 
SOUR1:DCO? 
>+3.0000E+00 
The offset for the current mode is set to +3 volts. 
SOURce[1]:SQUare:DCYCle 
Set
Query
Description 
Sets or queries the duty cycle for square waves 
only. The setting is remembered if the function 
mode is changed. The default duty cycle is 50%.  
Note 
The duty cycle of square waveforms depend on the 
frequency settings. 