106 
SOURce[1]:AMPlitude 
Set
Query
Description 
Sets the output amplitude for the 
SOURce[1]:FUNCtion command. The query 
command returns the current amplitude settings. 
Note 
The maximum and minimum amplitude depends 
on the output termination. The default amplitude 
for all functions is 100 mVpp (50Ω).  
The offset and amplitude are related by the 
following equation.  
|Voffset| < Vmax – Vpp/2 
The output amplitude can be affected by the 
function and unit chosen. Vpp and Vrms or dBm 
values may have different maximum values due to 
differences such as crest factor. For example, a 
5Vrms square wave will be adjusted to 3.536 Vrms 
for a sine wave. 
The amplitude units can be explicitly used each 
time the SOURce[1]:AMPlitude command is used. 
Alternatively, the VOLT:UNIT command can be 
used to set the amplitude units for all commands. 
Syntax 
SOURce[1]:AMPlitude {< amplitude> 
|MINimum|MAXimum} 
Example 
SOUR1:AMP MAX 
Sets the amplitude to the maximum for the current 
mode. 
Query Syntax 
SOURce[1]:AMPlitude? {MINimum|MAXimum} 
Return Parameter 
<NR3> 
Returns the amplitude for the current 
mode. 
Query Example 
SOUR1:AMP? MAX 
>+5.0000E+00 