User manual
    REMOTE INTERFACE 
103 
Output Commands
Unlike the Apply commands, the Output commands are low level 
commands to program the function generator.  
This section describes the low-level commands used to program the 
function generator. Even though the APPLy command is the easiest 
way to program the function generator, it lacks the ability to change 
individual parameters. The Output commands on the other hand 
can be used to set individual parameters, or those parameters that 
cannot be programmed with the Apply command. 
SOURce[1]:FUNCtion 
Set
Query
Description 
The FUNCtion command selects and outputs the 
selected output function. The User parameter 
outputs an arbitrary waveform previously set by 
the SOURce[1]:FUNC:USER command. 
The previously set frequency, amplitude and offset 
values are used automatically. 
Note 
If the function mode is changed and the current 
frequency setting is not supported by the new 
mode, the frequency setting will be altered to the 
next highest value.  
Vpp and Vrms or dBm amplitude values may have 
different maximum values due to differences such 
as crest factor. For example, if a 5Vrms square 
wave is changed to a sinewave, then the Vrms 
value is automatically adjusted to 3.536Vrms. 
The modulation and sweep modes can only be 
used with some of the basic waveforms. If a mode 
is not supported, the conflicting mode will be 
disabled. See the table below.  
Sine 
Square 
Ramp 
Noise 
ARB 
AM 
   
   
   
   
   
FM 
   
   
   
   
   