SYSTem:KLOCk[:STATe]
This command locks or unlocks the instrument front panel controls. The query
command returns “0” (OFF) or “1” (ON).
Group
System
Syntax
SYSTem:KLOCk[:STATe] {ON|OFF|<NR1>}
SYSTem:KLOCk[:STATe]?
Arguments
ON or <NR1>0 locks front panel controls.
OFF or <NR1>=0 unlocks front panel controls.
Returns
<NR1>
Examples
SYSTem:KLOCk:STATe ON
locks front panel controls.
SYSTem:PASSword:CDISable (No Query Form)
This command disables protected commands. The instrument security protection is
activated.
In the AFG3000 Series Arbitrary/Function Generators, no remote commands are
under the control of SYSTem:PASSword commands.
Group
System
Related Commands
SYSTem:PASSword[:CENable]
SYSTem:PASSword[:CENable]:STATe?
SYSTem:PASSword:NEW
Syntax
SYSTem:PASSword:CDISable <password>
Arguments
<password>::=<string> specifies current password. The string is case sensitive.
Examples
SYSTem:PASSword:CDISable <password>
activates the security protection.