[SOURce[1|2]]:PULSe:WIDTh
This command sets or queries the pulse width for the specified channel.
Pulse Width = Period × Duty Cycle / 100
The pulse width must be less than the period. The setting range is 0.001% to 
99.999% in terms of duty cycle.
AFG3011: 80 ns to 999.99 s
AFG3021B and AFG3022B: 30 ns to 999.99 s
AFG3101 and AFG3102: 8 ns to 999.99 s
AFG3251 and AFG3252: 4 ns to 999.99 s
Pulse Width ≤ Pulse Period – 0.8 × (Leading Edge Time + Trailing Edge Time)
Pulse Width ≥ 0.625 × (Leading Edge Time + Trailing Edge Time)
Group
Source
Related Commands
[SOURce[1|2]]:PULSe:DCYCle
Syntax
[SOURce[1|2]]:PULSe:WIDTh {<seconds>|MINimum|MAXimum}
[SOURce[1|2]]:PULSe:WIDTh?[MINimum|MAXimum]
Arguments
<seconds>::=<NRf>[<units>]
where
<NRf> is the pulse width.
<units>::=[ns | μs | ms | s]
Returns
<seconds>
Examples
SOURce1:PULSe:WIDTh 200ns
sets the CH 1 pulse width to 200 ns.