User guide
Page | 15
MRK-D-0022 V3.1 Aeroqual Series 200, 300 & 500 User Guide
4.1.2. Configuration
The Software can be used to set up the Monitor ID, Location Id, Alarm and Control Limits,
Logging Frequency, Units of Measurement, Output Sensor and Updating the Clock.
Click Monitor Setup (or Ctrl P)
The monitor setup dialog box will appear
Type in the required values or scroll up or down to set
the desired values.
Click the save button to synchronise the monitor
values with those on the PC.
In order to data log it is essential that the clock on the
monitor is set to the real time. This can be done on
the monitor (Section 4.4.6.) or via the PC by using the
“Update Clock” button in the configuration menu.
Note: If the temperature and humidity sensor is
connected an option to select temperature unit will
appear
4.2. Data Logging
The Series 500 monitor can log up to 8188 data points divided by the number of sensors being
logged. For example, if a triple sensor is connected it will log 8188/3 readings. Each sensor uses one
of the records each log period. Refer to Section 3.4.11 for instructions on how to log data on the
monitor itself. Alternatively, data can be logged directly to a PC.
4.2.1. PC Data Logging
Ensure that the USB cable is connected to the monitor and to the PC. Click File Start PC Data
Logging or click on the toolbar. The data will be logged directly to the database on the PC. No
data will be stored on the monitor memory in this mode of operation. However, if PC data logging is
stopped by clicking on the toolbar the data will start to be stored in the Series 500 monitor
memory.
4.2.2. Downloading Logged Data from Series 500
Logged data from the monitor memory can be downloaded by clicking File Download Logged Data
or by clicking on the toolbar. A data download progress bar will appear while the data is
downloading.
Note: The data will not be removed from the monitor memory. To delete the logged data from
the monitor memory select “Clear Log” on the display menu under Logging Setup.