User guide
Page | 12
MRK-D-0022 V3.1 Aeroqual Series 200, 300 & 500 User Guide
For compatibility purposes the database records for earlier versions of the S500 (v5.x) default
to monitor ID = 1 and location ID = 1. This cannot be changed.
Press “enter” to confirm the ID and return to the set up menu.
Note: In the Series 500 data logging mode only the Location ID will be logged. The monitor ID
is loaded to the database at the time of data upload (where it is logged).
3.4.7. Output Sensor (Series 300 & 500)
If a multi-sensor head is connected, the output sensor selection will determine which sensor
the alarm and control points and buzzer refer to
The output sensor can be found under “MONITOR SETUP”, scroll down in the monitor set up
menu and select “OUTPUT SENSOR” and select the relevant parameter
The 0-5v analogue signal will also operate on the selected sensor and measurement range
3.4.8. Alarm Points (Series 300 & 500)
The alarm points can be found under “MONITOR SETUP”, scroll down in the monitor set up
menu and select “ALARM POINTS”
Once the alarm set up menu has been entered “ALARM HI” will display
Press the “scroll up” and “scroll down” buttons to select the correct concentration
Press “enter” to confirm the change
Once the high alarm has been set “ALARM LO” will display
Press the “scroll up” and “scroll down” buttons to select the correct concentration
Press “enter” to confirm the change
Once the low alarm has been set “BUZZER” display will show. Here you can either enable or disable
the alarm. If the alarm is disabled a “P” will show on the right hand side of the alarm mute symbol on
the main display, highlighting that the alarm has been permanently muted.
If the concentration reaches above the high alarm point two flashing arrow symbols pointing
upwards will appear on the display, accompanied by a fast beeping sound.
If the concentration reaches above the low alarm point two flashing arrow symbols pointing
downwards will appear on the display, accompanied by a slow beeping sound.
Note: The alarm can be muted by pressing the mute button on the display screen. However,
this is only a temporary mute and will only mute the alarm during that specific alarm condition.
Once the concentration has moved back into the alarm limits the settings will reset. To turn of
the mute permanently follow instructions to disable the alarm.