Series 300, 305, 500 & 505 200Monitor User Guide 1
Quick Start Guide 1. Unpack monitor and check supplied components are correct Insert Battery (see section 3.0) Insert Sensor When new, plug in the power adaptor and turn on your monitor (with sensor head installed) for 24hrs. Note: The battery requires 15hrs from full discharge. The sensor requires a run in period. Thereafter, the normal warm up takes only 10 minutes. If your unit is new or unused for a long period, a warm up message may display for several hours, then will disappear.
Contents 1.0 Monitor Components 2.0 About Your Monitor 3.0 Assembly and Installation 3 4 5 3.1 Battery pack 3.2 Sensor warm up 3.3 Monitor and Sensor Versions 5 8 8 4.0 Operating Instructions 9 4.1 Turning the Monitor on and off 4.2 Operating under battery power 4.3 Sensor Head Failure 4.4 Using the Menus & Key Functions 4.5 Zero Calibration 4.6 Changing Unit of Measurement 4.7 Max Min Average Measurement Cycle 4.8 Alarm Mute (Perc only) 4.9 Alarm (Perc only) 4.
1.0 Monitor Components Series 200 Monitor The following components are supplied with the Series 200 Monitor: • • • • • Series 200 monitor base Gas sensor head 12 VDC 800 mA AC/DC adaptor User guide Battery pack – 9.6 V Please check that all these components have been supplied and contact your dealer or Aeroqual by email at: sales@aeroqual.com if any of the components are missing. Replacements Replacement batteries, sensor heads and other accessories can be purchased separately.
2.0 About Your Monitor The Aeroqual Series 200 Monitor has been specifically designed to incorporate Aeroqual’s in-depth knowledge of accurate ambient gas measurement. Different sensor heads are used for specific gases or depending upon whether high or low concentrations are to be measured. The all sensor heads are interchangeable on the same base unit. The following information details the operation and features of the monitor: 1.
3.0 Assembly and Installation Assembly The following needs to be completed before the Monitor is ready for use: • Insert the sensor head into the top of the monitor. The sensor head is keyed to ensure the head is inserted correctly. • If the monitor is a mains only unit it is ready for use. If the monitor is to be used as a portable unit with batteries then the batteries must be installed and charged prior to use. 3.1 Battery Pack • Remove the battery cover from the monitor.
Removing the Battery Pack • Remove the battery cover from the monitor. • Insert your thumb under the top left hand side of the battery pack and carefully lever the battery pack out from the retaining clips. • Unclip the connector from the monitor. • Replace the battery cover. Rechargeable Batteries The optional Ni-MH (Nickel Metal Hydride) battery pack is used to power the monitor as a portable unit.
Charging and Discharging the Batteries New Ni-MH batteries must be fully charged prior to use. A new NiMH battery's full performance is achieved only after a number of complete charge and discharge cycles. A fully charged battery will run for approximately 4-5 hours. (Note: Perc sensor heads require more power and will run for a shorter time) Charging Instructions 1) Plug in the adaptor 2) Turn the monitor on 3) Allow 15 hours for the battery to fully charge.
Installing the AC/DC adaptor • Connect the AC/DC adaptor to the mains supply. Ensure that the adaptor is correctly rated for the power supply. • Plug the adaptor into the monitor. The monitor can now be used as a mains unit. If the monitor is to be used as a portable unit the batteries will need to be charged. 3.2 Sensor warm up Prior to operation the monitor must be warmed up to burn off contaminants on the sensor. When the monitor is first switched on it will warm up for 3 minutes.
4.0 Operating Instructions The following instructions detail the operation and set up of the monitor. Power button Enter Key 4.1 Turning the monitor on and off To turn the monitor on, press & hold the power button until the screen activates. The monitor will turn on and the display will show “AEROQUAL MONITOR S200 / S205 Vx.x”,. After 5 seconds the display will change to indicate the type of sensor head. The monitor will warm up as described in “Warm up” in section 3.2.
Turning Standby Mode on and off To activate Standby Mode with the monitor on, press the power button once. This will stop operation of the monitor; however the sensor will be kept warm. This mode is used to conserve power between operations but keep the sensor ready to measure the gas concentration with only a short warm up phase as described in “Warm up” in section 3.2. In this mode the display will be as shown below: NH3 To return the monitor to operational mode press the power button once.
4.4 Using the Menus and Key Functions To enter the setup menu press appear: Normal Menu Special Perc Menu EXIT ZERO CAL CONC UNIT MIN MAX AV Press The following menus will EXIT ZERO CAL CONC UNIT MIN MAX AV ALARM MUTE ALARM ALARM LEVEL or to scroll to the desired menu. 4.5 Zero Calibration The “ZERO CAL” routine resets the zero point of the monitor. "ZERO CAL" must be carried out under carefully controlled conditions, as specified below.
4.6 Changing Unit Of Measurement (Concentration Unit) To enter the ‘CONC UNIT’ menu press will appear. . The unit display OZONE UNIT: ppm (or) mg/m3 Press to scroll between ‘ppm’ or ‘mg/m3’ Press to confirm the unit selection Press to scroll down to the next Menu choice 4.7 The Max/Min/Average Measurement Cycle Pressing starts the Min Max Av routine Press to scroll to ‘EXIT’ Press to Exit out of Menu Setup. The following display will appear:VOC O3 LOW . M I N 0.005 M A X 0.024 AV 0.
4.8 ALARM MUTE (for PERC only) Note: With “Alarm Mute ON” (enabled) the alarm condition will be muted only until the gas concentration drops below the alarm level – remember this is a temporary mute function. To enter the “ALARM MUTE” menu press screen will appear: - and the following ALARM MUTE OFF Press to toggle between “ALARM MUTE OFF” and “ALARM MUTE ON”. Press to confirm your choice and you will return to the main menu.
If you select “Alarm Permanent OFF” and exit the menu, the P will be displayed on the gas concentration symbol screen to indicate that the alarm has been permanently muted. Press to scroll down to the next menu choice. 4.10 ALARM LEVEL (for PERC only) To enter the “ALARM LEVEL” menu press ing screen will appear: - and the follow- ALARM LEVEL 50PPM Press 35PPM. to toggle between 50PPM, 100PPM, 150PPM and Press to confirm your choice and you will return to the main menu.
5.0 Troubleshooting Series 200/205 Fault Description Possible cause Remedy No power Lead connection broken Reconnect power lead Power supply failure Replace 12V DC power supply Battery flat Recharge battery S200 damaged Replace unit Sensor head damaged Replace sensor head Sensor failure when the sensor is new Insufficient warm up Run the sensor on full power for 24-48 hours.
Fault Description Possible cause Remedy Reading higher than expected in the presence of sensor gas Sensor correct Check calibration of gas generator.
6.0 Specifications Series 200 Monitor Base Measurement units PPM or mg/m3 Removable / replaceable sensor head Yes On-board alarm* Yes Alarm status displayed * Yes Low battery indication Yes Alarm mute function* Yes Sensor condition status Yes Stand-by mode Yes External signal for alarms & control None External signal functions None Analogue output None Power supply 12 VDC, 800 mA Rechargeable battery pack 9.
7.0 Drawings Depressions in the battery cover demarcating the screw-hole positions for fixing the unit to the wall.
8.0 Care and Maintenance Your Aeroqual Monitor is a product of superior design and quality and should be treated with care. When using your Aeroqual Monitor: • Keep it and all its parts and accessories out of the reach of small children. • Keep it dry. Avoid water and/or condensation as humidity and liquids containing minerals may corrode electronic circuits. • Do not use or store in dusty, dirty areas. • Do not store or turn off the monitor in temperatures below 10°C.
9.0 For Your Safety Read these simple guidelines. Ignoring these guidelines may be hazardous. USE SENSIBLY Use only as per this user guide. USE AEROQUAL APPROVED SERVICE Only approved service personnel must work on this product. ACCESSORIES Use only approved accessories. Do not connect incompatible products. CONNECTING TO OTHER DEVICES When connecting to any other device, read the appropriate user guide for detailed safety instructions. Do not connect incompatible products.
10.0 Statements of Compliance 1. The Aeroqual Series 200 Monitor and Remote Adaptor Kit comply with EN 50082-1:1997 The Aeroqual Series 200 Monitor and Remote Adaptor Kit comply with EN 50081-1:1992 The Aeroqual Series 200 Monitor and Remote Adaptor Kit comply with Part 15 of the FCC Rules.
11.0 Warranty Thank you for purchasing this Aeroqual product. To get maximum use of the features of your new product we recommend that you follow a few simple steps: Read the guidelines for safe and efficient use. Read all the terms and conditions of your Aeroqual Warranty. Save your original receipt. You will need it for warranty repair claims. Should your Aeroqual product need warranty service, you should return it to the dealer from whom it was purchased or contact Aeroqual.
Conditions • • • • • • • • • • The warranty is valid only if the original receipt issued to the original purchaser by the dealer, specifying the date of purchase, is presented with the product to be repaired or replaced. Aeroqual reserves the right to refuse warranty service if this information has been removed or changed after the original purchase of the product from the dealer.
12.0 Copyright Copyright Aeroqual Limited. All rights reserved. Reproduction, transfer, distribution or storage of part or all of the contents of this document in any form without the prior written permission of Aeroqual Limited is prohibited. “Aeroqual” and “Aeroqual Limited – Making the Invisible Visible” are registered trademarks of Aeroqual Limited. Other product and company names mentioned herein may also be trademarks or trade names. Aeroqual operates a policy of continuous development.
13.0 Guidelines on How to Measure Ozone The following information is presented to help users operate their Aeroqual monitor with an ozone sensor installed in the most effective and efficient manner. General Ozone is heavier than air and tends to sink. Thus detection • of leaks from ozone generating equipment should be performed at the most appropriate position. Ozone will react and decompose on surfaces such as walls, furniture etc.
14.0 Zero calibration additional Information A zero calibration menu routine is included in the Series 200, however it is important to understand when this should be used. All Aeroqual sensors are supplied pre-calibrated and zero calibration is performed in the factory in an atmosphere containing < 0.5 ppm hydrocarbons, 50% relative humidity and 25 oC. Zero baseline reading checks should be performed ideally under the same conditions.
15.0 Sensors and accessories Gases Ranges Ammonia (NH3) 0 - 100 ppm 0 - 1000 ppm Carbon monoxide (CO) 0 - 100 ppm 0 - 1000 ppm Carbon dioxide (CO2) 0 - 2000 ppm 0 - 5000 ppm 0 - 5% Hydrogen (H2) 0 - 5000 ppm Hydrogen sulphide (H2S) 0 - 10 ppm 0 - 50 ppm Methane (CH4) 0 - 10000 ppm Nitrogen dioxide (NO2) 0 - 0.200 ppm Non-methane hydrocarbon (NMHC) (Isobutylene standard) 0 - 25 ppm Ozone (O3) 0 - 0.150 ppm 0 - 0.