• Do not use components or attachments that are not
recommended by Aerogen. Refer to the Replacement
Parts and Accessories section in this manual (page 20).
• Never operate this product if any of the parts are not
working properly or have been damaged.
• Do not use the Aeroneb
®
Go Nebulizer in the presence
of a ammable anaesthetic mixture with air, oxygen or
nitrous oxide. As with any electrical equipment, take care
when using the device in a potentially oxygen enriched
environment.
• Do not use the Aeroneb
®
Go Nebulizer with anaesthetic
breathing systems or lung ventilator breathing systems.
• Do not place or drop this product into water or any liquid
when it is plugged into a power source.
• Keep this product and cords away from heated surfaces.
• Be sure the device has been properly cleaned before
use to avoid possible contamination.
• If after proper cleaning it takes longer than 12 minutes to
deliver 3 mL of normal saline (0.9%), the Nebulizer Unit
should be replaced.
• The user should ensure that exhaled medication is
directed away from bystanders.
• Do not microwave any parts.
• By connecting the AC/DC adapter to the Aeroneb
®
Go
Control Module, the Aeroneb
®
Go Control Module and
batteries are not being charged.
• As with any cables keep all cables tidy to avoid tripping
or strangulation hazards and take particular care around
children.