Instruction manual

Guidance and manufacturer’s declaration – electromagnetic immunity
The Aeroneb Go nebulizer system is intended for use in the electromagnetic
environment specied below. The customer or the user of the Aeroneb Go
should assure that it is used in such an environment.
Immunity
test
IEC/EN 60601
Test level
Compliance
level
Electromagnetic environment –
guidance
Field strengths from xed RF
transmitters, as determined by
an electromagnetic site survey, ª
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reection from structures, objects
and people.
ª Field strengths from xed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to xed RF
transmitters, an electromagnetic site survey should be considered. If the
measured eld strength in the location in which the Aeroneb Go is used
exceeds the applicable RF compliance level above, the Aeroneb Go should
be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orientating or relocating
the Aeroneb Go nebulizer system.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less
than [3]V/m