Instruction manual

Immunity test IEC/EN 60601
Test level
Compliance level Electromagnetic
environment –
guidance
Power
frequency
(50/60 Hz)
Magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency
magnetic elds
should be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
Note: Ut is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Aeroneb Go nebulizer system is intended for use in the electromagnetic
environment specied below. The customer or the user of the Aeroneb Go
should assure that it is used in such an environment.
Immunity
test
IEC/EN 60601
Test level
Compliance
level
Electromagnetic environment –
guidance
Conducted
RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 Vrms
150 kHz to
80 MHz
3 Vrms
80 MHz to
2.5 GHz
[3]V
[3]V
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Aeroneb Go, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
d = [1.17] √P
d = [1.17] √P... 80 MHz to 800 MHz
d = [2.33] √P... 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).