Instruction manual

18
PERFORMANCE:
Aerosol ow rate: > 0.3 mL/min
Average tested ow rate: 0.4 mL/min
Particle Size:
As measured with the Andersen Cascade Impactor:
- Specication Range: 1-5 μm.
- Average Tested: 3.1 μm.
As measured with the Marple 298 Cascade Impactor:
- Specication Range: 1.5-6.2 μm.
- Average Tested: 3.9 μm.
As per EN 13544-1:
-
Aerosol output rate: 0.25 mL/min
- Aerosol Output: 0.86 mL emitted of 2.0 mL dose
- Residual Volume: <0.1 mL for a 3 mL dose
Performance may vary depending on the type of drug and
nebulizer unit used. For additional information contact
Aerogen or your drug supplier.
The temperature of the medication will not rise more than
10°C (18°F) above ambient temperature during normal use.
For additional information contact Aerogen or your drug
supplier.