Operating instructions

ACCEPTANCE TESTING
5-42
Results table 5-33 RF broadband power meter (N-type input)
Frequency (MHz) Input
level (dBm)
Lower
limit (dBm)
Upper
limit (dBm)
Result
N-input
11 20 19.4 20.5 _______
100 20 19.4 20.5 _______
200 20 19.4 20.5 _______
300 20 19.4 20.5 _______
400 20 19.4 20.5 _______
500 20 19.4 20.5 _______
600 20 19.4 20.5 _______
700 20 19.4 20.5 _______
800 20 19.4 20.5 _______
900 20 19.4 20.5 _______
1000 20 19.4 20.5 _______
N in / N out
100 20 19.4 20.5 _______
100 7 6.4 7.5 _______
N input
100 7 6.4 7.5 _______
The upper and lower limits in the table above are calculated on the assumption that it was possible to set the exact
input level required. If it is not possible to achieve this, the limits will need to be calculated in proportion to the
power applied.