Operating instructions
ACCEPTANCE TESTING
5-40
Results table 5-27 AM measurement versus AM depth (1 kHz rate)
Depth (%) Lower
limit (%) †
Upper
limit (%) †
Result
5 3.75 6.25 _______
20 18 22 _______
50 46.5 53.5 _______
80 75 85 _______
95 89 100 _______
Results table 5-28 AM demodulation oscilloscope
Carrier
frequency
(MHz)
Depth (%) Lower
limit (%) †
Upper
limit (%) †
Result
100 60 54 66 _______
† The upper and lower limits in Results table 5-25, Results table 5-26, Results table 5-27, and Results
table 5-28 are calculated on the assumption that it was possible to set the exact level on the external
Modulation meter. If it is not possible to achieve this, the limits will need to be calculated for the
DVM readings obtained.