Operating instructions
ACCEPTANCE TESTING
5-34
Results table 5-8 Residual FM test points
UUT
RF generator
frequency (MHz)
Local oscillator
frequency
(MHz)
Limits
(Hz)
Result
1000 55.63889 12 _______
925 54.50000 12 _______
730 52.25000 12 _______
502 55.94444 12 _______
240 48.30000 12 _______
20 12 _______
Results table 5-9 AM versus carrier frequency (50% depth at 1 kHz rate)
Carrier
frequency (MHz)
Lower
limit (%)
Upper
limit (%)
Result (%)
1.5 46.5 53.5 _______
50 46.5 53.5 _______
100 46.5 53.5 _______
150 46.5 53.5 _______
200 46.5 53.5 _______
250 46.5 53.5 _______
300 46.5 53.5 _______
350 46.5 53.5 _______
400 46.5 53.5 _______
Results table 5-10 FM versus carrier frequency (at 10 kHz deviation 1 kHz rate)
Carrier
frequency (MHz)
Lower
limit (kHz)
Upper
limit (kHz)
Result
0.5 9.290 10.710 _______
10 9.290 10.710 _______
100 9.290 10.710 _______
200 9.290 10.710 _______
300 9.290 10.710 _______
400 9.290 10.710 _______
500 9.290 10.710 _______
600 9.290 10.710 _______
700 9.290 10.710 _______
800 9.290 10.710 _______
900 9.290 10.710 _______
925 9.290 10.710 _______
975 9.290 10.710 _______
1050 9.290 10.710 _______