Operating instructions
ACCEPTANCE TESTING
5-25
(8) Disconnect the external reference from the UUT and connect an external reference with an
accuracy of 1 part in 10
9
or better to the RF signal generator.
(9) Refer to Results table 5-31 on page 5-41. Set the RF signal generator to provide a frequency
of 1000 M
Hz at a level of 0 dBm.
(10) Check that the UUT indicates a frequency between 999.999750 MHz and
1000.000250 MHz. The test limits in this step are for guidance and assume that the internal
frequency standard has been recently adjusted. Ageing and stability have to be considered
when establishing the real test limits.