Operating instructions

ACCEPTANCE TESTING
5-7
BNC output mode
No claim is made on the output level accuracy of the BNC socket. The following is a functional
check to ensure correct internal operation.
(1) Connect the 6920 sensor to the BNC output, as shown in Fig. 5-2.
(2) Set the UUT to [Rx TEST], RF IN/OUT [SELECT] BNC output, [RF Gen],
[LEVEL] 21 dBm, [FREQ] 10 MHz. All modulation and noise measurements should be
switched OFF.
(3) Check that the power meter indicates 21 dBm ± 4 dB.
(4) Repeat step 3 for 100 MHz and then in 100 MHz steps up to and including 1000 MHz.
Attenuator accuracy
The following test will confirm that the attenuator performs to the published performance
specification. In the event of the receiver not being available, an alternative method to
functionally test the attenuator is also suggested.
Measuring
receiver
C2582
UUT
N TYPE
OUTPUT
Fig. 5-3 RF level set up
(1) Connect the test equipment as shown in Fig. 5-3.
(2) Set the UUT to [RF Gen], [ LEVEL], −21 dBm, [ FREQ] 2.5 MHz.
(3) Tune the receiver to the frequency set on the UUT RF Generator and measure the RF level.
(4) Decrement the output of the UUT in 10 dB steps down to an RF level of 121 dBm,
measuring the RF level at each step.
(5) Repeat steps 2 to 4 for frequencies of 500 and 1000 MHz.