Operating instructions
LOCAL OPERATION
3-96
Unlocking stores
Insert the card in the Service Monitor and access the MEMORY CARD FORMAT screen
as described above. The legend 'Locked' will be shown against the [Change Code] and
[Lock Stores] keys.
Press the [Lock Stores] key, which will cause the message Enter Code to be displayed.
Key in the four figure code number and press the [ENTER] key. If the correct code
number has been entered the 'Locked' legend will no longer be displayed.
The inverse video legend Lock Stores 10 to 'n' is displayed ('n' being the number of the
highest store location protected).
Key in any store location address lower than 10 and press the [ENTER] key.
Changing the range locking
To change the range of the locked portion, proceed as above but key in the address of the
last location of the new portion to be protected and press the [ENTER] key.
Changing the locking code
To change the locking code, use the following procedure:-
Insert the card and access the MEMORY CARD FORMAT screen.
Press the [Change Code] key, which will cause the message Enter Code to be displayed.
Key in the existing code number and press the [ENTER] key. The 'locked' and the
ENTER CODE legends will no longer be shown.
Press the [Change Code] key. The inverse video legend 'New Card Code' is displayed.
Key in a new four figure number and press the [ENTER] key.
The new code will apply to the card but it will not be protected in the Service Monitor until the
next power-up.
Storing on memory cards
Storing screen bitmaps
The facility to store a screen bitmap allows any active test screen to be held and saved. This
includes oscilloscope traces, spectrum analyzer displays and harmonic analysis displays.
Screen bitmaps are stored from the DISPLAY HOLD STORE screen.
To store a screen proceed as follows:-
Press the [DISPLAY HOLD] key, as described under 'DISPLAY HOLD STORE, Storing
results', above, until the DISPLAY HOLD STORE screen is shown.
Press the [Store Screen] key. The inverse video message, 'Store Screen To Store No.' is
displayed.
Key in the required two digit address.
The location address list will show the contents of the location as 'Screen'.
Storing systems data
Settings and results of SYSTEM tests, and user defined test programs, can only be stored on a
memory card when the Service Monitor is operating in a system option. The procedures are
described in the operating manual supplement applicable to the system. A summary of operation
is:
A memory card location which contains SYSTEM data will show 'Syst Set' or 'Syst Res' or
'BASIC' on the location address list.
To store 'SYSTEMS' data, press the [MEM] key. The STORE/RECALL screen will be displayed.
A [Store Sys Par] key is shown, in addition to those on the non-system screen.
Pressing this key will show a new soft key menu with the three keys [Store Sys Set],
[Store sys Res] and [Store BASIC]. Pressing the appropriate key will store the current system
parameter settings, the last test results from the current system or the user defined test program if
one is present.
Titling, protecting or deleting can be used as for non-system operation.