User's Manual

4) Rinse the turbine by immerging
it in clean water (not hot).
5) Shake off the excess water
from the turbine and leave it to
dry, standing it vertically on a
dry surface.
6) Once the turbine has been
cleaned insert the turbine tube
in its place. To correctly insert
the turbine push it to the end
and turn it clockwise until
reaching the wedge which
ensures that the tube has been
blocked inside the plastic
casing.
CAUTION: RISK OF
CROSS-CONTAMINATION
To avoid cross contamination,
Aerobit Turbines are marked as
single Patient Use accessories.
However, to avoid self-
contamination, Aerobit Turbines
should be properly cleaned before
every use.
The Aerobit Reusable Turbine has a
paper Mouthpiece attached at one
end; a new unused mouthpiece must
be used for each set of tests. Used
mouthpieces must be discarded as
household wastes.
The correct functioning of the re-
usable turbine can only be
guaranteed if it has been cleaned in
the correct manner and is free from
foreign bodies which could alter its
movement. The cleaning of the
turbine should be performed
according to the instructions
contained in the User’s Manual.
PeakFlow
+
Tests
Parameters
FEV1(Forced Expiratory Volume in 1
second) The volume of air (l) A
forced maximal expiratory effort
after a full inspiration.
PEF (Peak Expiratory Flow) The
maximum flow rate (l/s) during a
forced maximal expiratory effort
after a full inspiration.
The final results displayed in the
application are always the best