User's Manual

• Never holding the turbine under a
jet of water or air and never let it
come into contact with high
temperature fluids.
•Not allowing dust or foreign bodies
to enter the turbine sensor, in order
to avoid incorrect functioning and
possible damage. The presence of
any impurities such as hair, sputum,
threads etc. within the body of the
turbine sensor may seriously
compromise the accuracy of the
measurements.
• Avoiding environmental
contamination by cleaning waste
products, the user must adhere to all
relevant regulations. The Reusable
turbine must be properly disinfected
before each test.
It is highly recommended to use
Aerobit Reusable Turbine and
Mouthpieces only, usage of third-
party turbines or mouthpieces could
seriously impact the accuracy of the
readings and safety of the patient.
Cleaning the Aerobit
Reusable Turbine
To clean the reusable turbine, please
refer to the following procedure: -
1) Remove the Turbine from its
compartment on the
PeakFlow
+
by turning it anti-
clockwise and pressing lightly.
It can be helpful to push it
gently from underneath with
one finger.
2) Immerse the turbine in the
recommended cold detergent
solution and move it within the
liquid to remove any impurities
which may be deposited inside.
3) Leave the turbine immersed for
the time specified in the
instruction of the solution. To
avoid any kind of damage to
the reusable turbine please do
not use any alcoholic or oily
substances, do not immerge
the turbine in hot water or hot
solution. Do not put the turbine
under a direct jet of water or
other liquid. If no detergent
solution is available, clean the
turbine in clean water.