User's Manual
It is highly recommended to use
Aerobit Reusable Turbine and
Mouthpieces only, usage of third-
party turbines or mouthpieces could
seriously impact the accuracy of the
readings and safety of the patient.
Cleaning the Aerobit
Reusable Turbine
To clean the reusable turbine, please
refer to the following procedure: -
1) Remove the Turbine from its
compartment on the SpiroLite
by turning it anti-clockwise and
pressing lightly. It can be
helpful to push it gently from
underneath with one finger.
2) Immerse the turbine in the
recommended cold detergent
solution and move it within the
liquid to remove any impurities
which may be deposited inside.
3) Leave the turbine immersed for
the time specified in the
instruction of the solution. To
avoid any kind of damage to
the reusable turbine please do
not use any alcoholic or oily
substances, do not immerge
the turbine in hot water or hot
solution. Do not put the turbine
under a direct jet of water or
other liquid. If no detergent
solution is available, clean the
turbine in clean water.
4) Rinse the turbine by immerging
it in clean water (not hot).
5) Shake off the excess water
from the turbine and leave it to
dry, standing it vertically on a
dry surface.
6) Once the turbine has been
cleaned insert the turbine tube
in its place. To correctly insert
the turbine push it to the end
and turn it clockwise until
reaching the wedge which
ensures that the tube has been
blocked inside the plastic
casing.
CAUTION: RISK OF
CROSS-CONTAMINATION