Ideal for your industrial process applications, such as measure-
ment, dosing or environmental monitoring, the L452 is equipped
with 2 independent inputs. It allows you to connect various types
of sensors to the input in order to record varied data:
• 4-20 mA modules
• 0-10 V modules
These modules can also be used to record various data such as:
• temperature
• humidity
• air speed
For each of the channels, a real-time measurement is displayed on
the LCD screen. The L452 communicates via a USB input and the
Bluetooth function for configuration and data transfer onto a PC.
Specially designed for Chauvin Arnoux
®
Data Loggers, the control
panel for the data analysis software is a module fully integrated in
the Dataview
®
software, which has already proven its value.
The instrument can also be configured using its front panel, with a
very simple user interface. The menus are multilingual. Users can
lock the keypad with the PC software.
Alarms can be set on the various measurements. In counter mode, it
displays the measurement results in the unit selected (Wh, m.s, etc.).
The measurement campaign, which is programmable, can be
launched very simply using the L452’s keypad.
Its ergonomics makes it even simpler to use: magnetized rear panel,
Multifix system, desktop stand. It operates on rechargeable batteries
or the mains supply.
Measurement range Resolution Accuracy (% of reading) Sampling rate
DC current 4 to 20mA 0.01 mA 0.05 mA (0.25 %) 5 S/s
DC voltage
100 mV 0.1 mV 0.1 mV (0.5 %)
5 S/s1 V 1 mV 1 mV (0.5 %)
10 V 10 mV 10 mV (0.5 %)
Pulse n/a 1 ms n/a n/a
Digital n/a 1 ms 1 s
(recording for 1 month max.) n/a
Pulse voltage 3.3 V (with 1,000,000Ω pull-up)
Battery life
200 ms acquisition, display on: 18 days
200 ms acquisition, display off: 36 days
1 min. acquisition, display off: 270 days
Power supply 110 to 240 V (50/60Hz) – External: via USB connector – Internal: rechargeable NiMH batteries, 2.4 V (2 x 1.2 V)
Recording modes Start/Stop (stop when memory full or when campaign end date is reached)
Control Local mode (multi-directional keypad on front panel) – Remote mode (control via PC)
Recording duration 10 minutes to 1 year, modifiable
Examples
2 channels at @ 200 ms: 19 days
2 channels @ 1 min: > 1 year (theoretically)
Acquisition interval 200 ms to 1 hour
Communication Bluetooth 2.1, Class 1, USB 2.0
Dimensions Thickness: 32.4 mm – Width: 65.5 mm – Length: 125 mm (137.5 mm with screw connector)
Weight 206 g
Display LCD 128 x 64 pixels
Measurement terminal strip 6 screw terminals
Temperature Operation: 0 to 50 °C
Protection IP42 (IP20 terminal)
Electrical protection IEC 61010-1 Ed. 3 and IEC 61010-2-030 Ed. 1
State at delivery:
The L452 Data Logger is delivered with:
• User’s manual
• µUSB power cable
References and accessories:
L452 P01157201
DataView
®
software P01102095
µUSB power cable P01102148
Wall-mount P01651024
Multifix mounting adapter P01102100Z
Screw connector kit (x 5) P01295489