Operating instructions
31
3.1.9 UPLOAD NEW LBA-2 FIRMWARE
This function enables, if required, current LBA-2 firmware to be uploaded (if available).
The software update can be copied by PC or Notebook on to the SD card of the corresponding
LBA-2. Do not open up a new folder on the SD card, but copy the file in the SD card’s root directo-
ry and if necessary, replace the existing file. As soon as the SD card is inserted in the LBA-2, and the
latter is inserted into the active Thyro-P, the firmware update will load automatically. A progress bar
will appear that shows the remaining waiting time.
3.1.10 LANGUAGES
In the standard version of LBA-2 are the following languages available: German, English, Chinese,
French, Swedish, Czech, Turkish and Spanish.
The languages can be selected in the LBA-2 menu, see table LBA-2 MENU STRUCTURE.
On request further languages can be implemented.
3.2 LBA-2 TOOL
The LBA-2 Tool (free of charge) offers users to store data on the SD card and to analyze the accor-
ding process data in combination with data logger entries via PC or Notebook. First the LBA-2 Tool
program must be installed on your PC or Notebook. The LBA-2 Tool can be downloaded from the
AEI website:
www.advanced-energy.com
All downloaded data have to be copied into a directory. LBA-2 Tool.exe starts up the application.
Then the LOG files, in the folder LOG, can be selected and opened by the LBA-2 Tool. To use the
files with LBA-2 Tool, the LOG files can be saved locally on the PC or another medium.