Step 3: Adjust the registers until they fall between the input voltage
from the standard voltage reference and the receiving voltage
reflectected in the Manual A/D Calibration tab.
D/A channel Manual-Calibration
Step 1: Click the Manual D/A Calibration tab to show the D/A
channel manual calibration panel. Two D/A channels are
individually calibrated . Before calibrating, output desired
voltage from the D/A channels and measure it through an
external precision multimeter.
Step 2: For example, choose channel 0; select the Range and select
the wished output voltage code or value from the radio
buttons (Fig. 5-17 and Fig. 5-18).
Figure 5-17 & Figure 5-18: Selecting D/A Range and Choosing
Output Voltage