Specifications

71
5.3 PCI-1716/1716L Calibration
A calibration utility, AutoCali, is included on the companion
CD-ROM :
AutoCali.EXE PCI-1716/1716L calibration utility
This calibration utility is designed for the Microsoft©Windows™
environment. Access this program from the default location:
\Program Files\ADVANTECH\ADSAPI\Utilities\PCI1716
VR Assignment
There is one variable resistor (VR1) on the PCI-1716/1716L to adjust
the accurate reference voltage on the PCI-1716/1716L. We have
provided a test point (See TP4 in Figure 5-3) for you to check the
reference voltage on board. Before you start to calibrate A/D and
D/A channels , please adjust VR1 until the reference voltage on TP4
has reached +5.0000 V. Figure 5-3 shows the locations of VR1 and
TP4.
Fig. 5-3 PCI-1716/1716L VR assignment