62
voltage source.
Note:
Before you calibrate the A/D or D/A function, you must turn on the power
at least 15 minutes to make sure the DAS card getting stable.
5.1 PCI-1710/1710L/1710HG/1710HGL Calibration
Two calibration programs are included on the companion CD-ROM :
ADCAL.EXE A/D calibration program
DACAL.EXE D/A calibration program
(only for PCI-1710/1710HG)
These calibration programs are designed only for the DOS
environment. Access these programs from the default location:
\Program Files\ADVANTECH\ADSAPI\Utilities\PCI1710
VR Assignment
There are five variable resistors (VRs) on the PCI-1710/1710HG
card and three variable resistors (VRs) on the PCI-1710L/1710HGL
card. These variable resistors are to facilitate accurate adjustments
for all A/D and D/A channels. Please refer to the following two
figures for the VR positions.