D.3 Mode C waveform
Hardware-Triggered Strobe
Mode C is identical to Mode A, except that counting will not begin until
a Gate edge is applied to the armed counter, the counter must be armed
before application of the triggering Gate edge; Gate edges applied to a
disarmed counter are disregarded.
The counter will start counting on the first source edge after the
triggering Gate edge and will continue counting until TC. At TC, the
counter will reload from the Load register and automatically disarm
itself. Counting will then remain inhibited until a new ARM command
and a new Gate edge are applied in that order.
Note that after application of a triggering Gate edge, The Gate input will
be disregarded for the remainder of the count cycle. This differs from
Mode B, where the Gate can be modulated throughout the count cycle to
stop and start the counter.
SOURCE
WR
ARM
COMMAND
COUNT
VALUE
TC OUTPUT
TC TOGGLED
OUTPUT
GATE
X
L L-1 L-2 2 1 0 L
Mode C Waveforms