Operation Manual

Recommended Naída CI Operating and Storage Temperatures
Operating and Storage Temperature Ranges
Approved Power Sources
Advanced Bionics has verified the Naída CI to function with the power sources listed below;
behavior when using power sources other than those listed here cannot be guaranteed.
Approved Naída CI Power Sources
Approved Naída CI Power Supplies
Troubleshooting the Naída CI
If no sound is heard or the red LED is flashing once per second on the Naída CI processor LED
when the headpiece is in place over the implant, try the following troubleshooting steps:
Visually inspect the cable for any damage or breakage and verify it is firmly attached to
the headpiece and sound processor.
Remove any materials (hat, scarf, headband, etc.) that may be covering the microphone.
Unaided listeners may perform a listening check of the microphones as described in the
Listening to the Microphone and Auxiliary Input section of the User Guide.
Naída CI Sound Processor User Guide 53
Power Supply Rated Voltage Type
PowerCel
Zn-Air
Rated Energy
Programming Interface 5.0 V 1,250 mWDC
Condition Minimum Maximum
Operating temperature 0°C (32°F) 45°C (115°F)
Storage temperature
-20°C (-4°F)
55°C (131°F)
Power Source Model Number Capacity (mAhr)
CI-5511-XXX
110
PowerCel
CI-5517-XXX 170
CI-5523-XXX 230
PowerOne p675
Zn-Air
CI-6101
Programming Interface
(used during fitting, only)
1,140
Not Applicable
2.8 V
3.7 V Li-ion (Rechargeable) 0.4 Wh – 0.9 Wh
Zn-Air (Disposable) 1,240 mWh
1
2
3