Use Instructions
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
As with the operation of
other electronic
devices, precaution
should be taken to not
generate ESD.
Power frequency (50/60
Hz) Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz 3
3 V/m
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the Naída CI,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance:
d = 1.2√P < 800 MHz
d = 2.3√P ≥ 800 MHz
where P is the
maximum output
power rating of the
transmitter in watts (W)
according to the
transmitter
manufacturer and d is
the recommended
separation distance in
meters (m). Field
strengths from fixed RF
transmitters, as
determined by an
electromagnetic site