Manual

6-34 ADS Portable FlowShark Pulse Manual
Completed screen with correct settings
Example:
The measurement location can tend to experience backwater conditions and,
therefore, negative values must be recorded. The analog output signal for this
scenario must be set to floating.
As a result, configure the measurement span so that the unit will output a V signal for
flow = 0 reading in the middle of the measurement span.
Example:
0 V = -100 l/s (-2.28 mgd)
10 V = 100 l/s (2.28 mgd)
For this example, the signal output for a zero (0) reading would be 5 volts. Backwater
will cause the analog signal to decrease; positive flow will cause the analog signal to
increase.
The analog output will be updated during the measurement interval
.
While the unit is between two consecutive measurement cycles (in
sleep mode), the voltage value will be sustained based on the latest
value.
PAR: Digital Outputs
Submenu for digital outputs