6-22 ADS FlowShark IS Manual
Attempts Enter the number of times you want your local
PC to request data from the monitor following a failed
attempt while the monitor is still on line.
Low Temperature Enter the temperature reading from
the monitor, ultrasonic depth sensor, and pressure depth
sensor below which you want Profile to provide
notification.
High Tempera tu re Enter the temperature reading from
the monitor, ultrasonic depth sensor, and pressure depth
sensor above which you want Profile to provide
notification.
Log Communications Select this checkbox to record
all communication activities with the monitor.
Status on C o n n ect Select this checkbox to view details
concerning the current state of the monitor once monitor
communication has been established.
Low Battery 1502, 1506, 3500, 4000 (These
parameters do not apply to the FlowShark IS flow
monitor.)
Low Modem Battery TCP/IP Enter the voltage from
the battery inside the EMU powering the wireless modem
below which you want Profile to provide notification
concerning the wireless communication unit, when
applicable. ADS recommends leaving the default battery
setting unchanged.
Low Signal Strength TCP/IP Enter the signal strength
(in dBm) below which you want Profile to provide
notification concerning the wire communication unit, when
applicable. ADS recommends using -95 dBm for the low
signal strength.
Low Battery FlowShark (This parameter does not
apply to the FlowShark IS monitor.)
Low Battery FlowShark I S Internal Enter the
voltage, supplied either by the battery inside the monitor or