User Manual

6-2 ADS FlowShark Pulse O&M Manual
Transmitter Installation and Connection
General
When selecting a mounting location for the transmitter, avoid locations exhibiting or
near the following conditions:
z Direct sunlight (use weatherproof cover, if necessary)
z Heat-emitting objects (maximum ambient temperature of 104°F (40°C))
z Objects generating strong electromagnetic fields (such as frequency
converters or electric motors that consume significant power)
z Corrosive chemicals or gas
z Mechanical shock
z Walkways or heavy traffic areas
z Vibration
z Radioactive emissions
To fasten the wall mount enclosure, use one of the following fastener configurations
based on the composition of the mounting structure:
z Four M5 machine screws of suitable length with corresponding nuts and
shims. These screws must penetrate the wall at least 1.575 inches (40 mm).
z Four wood screws with a minimum diameter of 0.177 inches (4.5 mm).
These screws must penetrate the appropriate dowels at least 1.97 inches (50
mm).
The clear door on the measurement transmitter has a protective film to prevent
potential damage during transport and scratches during assembly. Remove this film
immediately after assembly.
Exposing the film to direct solar radiation for a long period will
make it more difficult to remove the film.
Clean the door with mineral spirits or car polish, when necessary. If these cleansers
are not effective, order a new front door from ADS.
The FlowShark Pulse Model 10/20 transmitter does not support
passive OCS sensors (older models).